Parameters:
<Unit>
POWer | VOLTage
Power or voltage units
*RST:
POWer
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'b1'
Create channel 4 and a trace named
Ch4Tr1
to measure the
wave quantity b
1
. The trace becomes the active trace in channel
4.
CALC4:FORM:WQUT VOLT
Select voltage units for the created trace (identified by the suffix
4).
Manual operation:
CALCulate<Chn>:GDAPerture:SCOunt
<Steps>
Defines an aperture for the calculation of the group delay as an integer number of fre-
quency sweep steps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Steps>
Number of steps
Range:
1 to 10000
*RST:
10
Example:
*RST; :CALC:FORM GDEL
Select group delay calculation for the active trace.
CALC:GDAP:SCO 15
Select an aperture of 15 steps.
Manual operation:
CALCulate<Chn>:IAVerage:MODE
<Mode>
Selects the quantities to be averaged if infinite averaging is enabled
(
CALCulate<Chn>:IAVerage[:STATe] ON
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Mode>
MAGPhase | REIMag
Selects the averaging mode.
MAGPhase
Averaging of magnitude and phase (default) of the complex
trace value
REIMag
Averaging of real and imaginary part of the complex trace value
SCPI command reference
www.allice.de