Concepts and features
R&S
®
ZND
183
User Manual 1173.9557.02 ─ 61
In the limit case, where the length of the transmission line is considered to be "almost
zero", the loss is considered constant:
Otherwise, if the loss at DC and one additional frequency
f
1
is known (or measured),
the loss at frequency f is approximated by:
If in addition the loss at a second frequency
f
2
is known (or measured), then the loss
can be approximated by:
In practice, the frequency-dependent part ist often dominant, so that
Loss
DC
can be set
to zero. Experimentally, the loss value at DC can be determined in a separate mea-
surement at a very low frequency (f --> 0).
4.6.1.3
Auto Length
The "Auto Length" function adds an electrical length offset to the active trace's receive
port, such that the residual delay of the active trace (defined as the negative derivative
of the phase response) is minimized across a configurable frequency range. If "Delay"
is the selected trace format, the entire trace is shifted in vertical direction and centered
on zero. In phase format, the "Auto Length" corrected trace shows the deviation from
linear phase.
Length and delay measurement, related settings
"Auto Length" is suited for length and delay measurements on transmission lines.
1. Connect a (non-dispersive) cable to a single analyzer port no. n and measure the
reflection factor S
nn
.
2. In the CHANNEL – [OFFSET
EMBED] > "Offset" softtool tab, select "Auto Length".
The delay is displayed in the "Delay" field, the cable length (depending on the
"Velocity Factor") in the "Mech. Length" field.
It is also possible to determine cable lengths using a transmission measurement. Note
that "Auto Length" always provides the
single
cable length and the delay for propaga-
tion in
one
direction.
The analyzer provides alternative ways for delay measurements:
1. Measure the reflection factor and select TRACE – [FORMAT] > "Delay".
Offset parameters and de-/embedding
www.allice.de
Allice Messtechnik GmbH