Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
[SENSe<Ch>:]CORRection:CONNection
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete
Cal Kit
Selects a cal kit for the connector at each selected physical port. The drop-down list
contains all available calibration kits for the selected connector type.
The assignment of a calibration kit to a connector type must be the same for all physi-
cal ports: If a calibration kit is changed, the R&S
ZND automatically assigns the new kit
to all ports with the same connector type.
Use "Import Cal Kit..." to add new kits to the list.
Remote command:
[SENSe:]CORRection:CKIT:SELect
Same Connector All Ports / Same Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some
multi-port calibration types, the port connector types must be equal, e.g. because they
require a Through standard with known characteristics.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Kit" dialog that allows you to import a cal kit file. For
background information, see
Chapter 4.5.2.4, "Cal kit Files"
Remote command:
Back
Go back to
.
Start
Start
.
Step 3: Calibration
Allows you to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration type. On
"Apply", the R&S
ZND calculates the system error correction (error terms) from the
measurement data of the standards and applies the result to the active channel.