Command reference
R&S
®
ZND
978
User Manual 1173.9557.02 ─ 61
Setting parameters:
<Activate>
ONCE
Applies the auto length function.
Example:
*RST; :CORR:EDEL:AUTO ONCE
Reset the instrument and apply the auto length function to the
default trace (Trc1 in channel 1).
Usage:
Setting only
Manual operation:
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DIELectric
<Permittivity>
Defines the permittivity for the offset correction at test port
<PhyPt>
.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<Permittivity>
Permittivity
Range:
1 to +1E+6
*RST:
1.00062
Example:
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
"Permittivity / Velocity Factor"
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DISTance
<MechLength>
Defines the offset parameter for test port
<PhyPt>
as a mechanical length.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<MechLength>
Mechanical length
Range:
-3.402038 m to +3.4028038 m.
*RST:
0 m
Default unit: m
Example:
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
"Delay / Electrical Length / Mech. Length"
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
<ElecLength>
Defines the offset parameter for test port
<PhyPt>
as an electrical length.
SCPI command reference
www.allice.de