
Command reference
R&S
®
ZND
960
User Manual 1173.9557.02 ─ 61
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected THRough, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected SHORt, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 2
:SENSe1:CORRection:COLLect:DISCard
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
<StandardType>,
<TestPort1>[, <TestPort2>]...
Starts a fixture compensation sweep in order to acquire measurement data for a test
fixture that has its inner conductor terminated with the selected standards.
Suffix:
<Ch>
.
Channel number of the corrected channel
Setting parameters:
<StandardType>
OPEN | SHORt
Terminating standard type: open or short.
<TestPort1>
<TestPort2>
Test port numbers. For a fixture compensation, n arbitrary (not
necessarily consecutive) port numbers must be specified.
*RST:
-
SCPI command reference
www.allice.de
Allice Messtechnik GmbH