GUI reference
R&S
®
ZND
516
User Manual 1173.9557.02 ─ 61
Windows Explorer
Opens the Windows Explorer and shows you the contents of the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\External Tools
appli-
cation shortcut directory.
5.15.2
Distance to Fault Application
With option K3, the R&S
ZND can locate faults and discontinuities on cables and trans-
mission lines. The measured S
ii
-parameter trace is mathematically converted to the
time domain, and represented in the distance domain. The x-axis of the DtF trace
(derived from S
ii
) shows the distance from the calibrated reference plane; faults appear
as peaks above a certain level.
If option R&S
ZND-K3 is not installed on the R&S
ZND, the "Distance to Fault" applica-
tion is not available.
To achieve the best distance resolution, set the sweep span as large as possible and
use the
to adjust the number of sweep points.
Background information
Refer to
Chapter 4.7.6, "Distance-to-fault measurements"
5.15.2.1
Controls on the Distance to Fault tab
The controls on the "Distance to Fault" tab allow you to enable and configure a stan-
dard Distance to Fault (DtF) measurement.