
Command reference
R&S
®
ZND
1019
User Manual 1173.9557.02 ─ 61
The frequency range is listed in
Note:
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
is equivalent to
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
. Source and receiver frequency are
always equal; the four commands overwrite each other.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency.
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
7.3.15.2
SOURce:GROup...
The
SOURce:GROup...
commands define the group of (un-)used logical ports.
........................................................................................ 1019
.............................................................................. 1020
............................................................................. 1020
............................................................................1020
SOURce<Ch>:GROup<Grp>
<arg0>, <arg1>
In "set direction" this command defines the set of active logical ports (the "master
group") as the continuous range from
<log_port1>
to
<log_port2>
, i.e. logical
ports outside this range are disabled.
In "get direction" it returns the minimum and maximum enabled logical port.
See
on how to define logical ports.
Suffix:
<Ch>
.
Channel number.
<Grp>
Port group number, must be 1 (or omitted).
Parameters:
<arg0>
<arg1>
First and last logical port number in the port group. Must be
omitted if the command is used as a query.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH