Concepts and features
R&S
®
ZND
208
User Manual 1173.9557.02 ─ 61
The rectangular "No Profiling (Rectangle) "windows minimize numerical inaccuracies
near the boundaries of the measured frequency span. In the limit where the effect of
the time gate vanishes (e.g. a gate of type "Notch" and a very small width), the time
gated trace is equal to the original measured trace.
4.7.5.6
Time domain S
VSWR
measurements
The vector network analyzer R&S
ZND with time domain option K2 supports TD site
VSWR measurements.
For the frequency range 1
GHz to 18
GHz, this method of S
VSWR
measurement is pro-
posed in standard ANSI
C63.25.
EMC Test Site Validation
EMC test sites for radiated emission measurements rely on free-space conditions to
minimize the influence of reflections on the received signal. Practically, near free-space
conditions are achieved by shielded enclosures fully lined with RF absorbing material.
Site validation
determines deviations from free-space conditions that must meet the
acceptance criterion for making EMC compliance measurements in a FAR (= fully
anechoic room). It is performed by measuring the
site voltage standing-wave ratio
(S
VSWR
), which is the ratio of maximum received signal to minimum received signal,
caused by interference between direct (intended) and reflected signals.
Optional extensions and accessories
www.allice.de
Allice Messtechnik GmbH