GUI reference
R&S
®
ZND
424
User Manual 1173.9557.02 ─ 61
The properties of the selected characterization are shown below the list. "Delete" dele-
tes the selected characterization file; "Start Characterization..." opens the
to create a characterization.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Characterization wizard
The "Characterization" wizard guides you through the automatic characterization of a
calibration unit.
Access:
> "Start Characterization..."
The guided characterization consists of the following steps:
1.
"Characterization":
Select the characterized ports and cal unit standards to ini-
tiate the characterization sweeps.
2.
"Save Characterization Data":
Save the characterization data to the calibration
unit.
Step 1: Characterization
Selects the calibration type and the characterized cal unit ports and initiates the neces-
sary characterization sweeps.
Characterization procedure
To acquire accurate characterization data, the test setup must be properly calibrated
before you start the characterization wizard. Use the calibration type that you wish to
perform with your new cal unit characterization; see
Chapter 4.5.5.3, "Characterization
Cal softtool