7
Impella LD
®
Circulatory Support System
15.
Press the
FLOW CONTROL
soft button to increase the P-level to P-9 to
confirm correct and stable placement. Verify placement with TEE and the
placement signal screen. Reposition the catheter as necessary.
Note:
When the flow rate is increased to maximum flow, the Impella
LD Catheter has a tendency to be drawn into the ventricle. Check
positioning at maximum flow to ensure proper placement throughout
the P-level setting range
16.
Position the front silicone plug as close as possible above the aorta.
Secure the silicone plug to the graft using a penetrating suture
ligature.
Note:
When securing the silicone plug to the graft, ensure that the
penetrating suture does NOT go all the way through the silicone plug
and damage the Impella LD Catheter inside the plug.
Transesophageal echocardiogram (TEE)
illustration showing correct Impella LD
®
Catheter position
Impella LD
®
Catheter
inlet area
Impella LD
®
Catheter
outlet area
RV
LV
LA
Aortic valve
Penetrating
Suture
Ligature
Rear Silicone Plug
Front Silicone Plug
Transesophageal echocardiogram (TEE)
illustration showing correct Impella LD
®
Catheter position
Impella LD
®
Catheter
inlet area
Impella LD
®
Catheter
outlet area
RV
LV
LA
Aortic valve
Penetrating
Suture
Ligature
Rear Silicone Plug
Front Silicone Plug
17.
Clear the vascular graft of excess blood and resecure the rear silicone
plug to the end of the graft.
18.
After achieving correct and stable placement, decrease the P-level to the
desired level for support.
Note:
After implantation of the catheter (and until explant), ACT should
be maintained at 160 to 180 seconds.
19.
Select the lowest P-level that will enable you to achieve the flow rate
necessary for patient support. You can select one of ten P-levels (P-0 to
P-9) for the Impella LD Catheter
P-LEVELS
You can select one of ten P-levels (P-0 to P-9) for the Impella LD
®
Catheter.
Select the lowest P-level that will enable you to achieve the flow rate
necessary for patient support.
P-Level
*Flow Rate (L/min)
P-0
0.0
P-1
0.0 – 1.4
P-2
0.5 – 2.6
P-3
0.5 – 3.1
P-4
0.9 – 3.4
P-5
1.4 – 3.7
P-6
1.8 – 4.0
P-7
2.6 – 4.4
P-8
3.4 – 4.7
P-9
4.2 – 5.3
*Flow rate can vary due to suction or incorrect positioning.
PURGE CASSETTE PROCEDURES
CHANGE CASSETTE & BAG
Follow these steps to change both the purge cassette and purge fluid:
1.
Press
PURGE MENU
and select “Change Cassette and Bag” from the
menu.
2.
Select
START
to begin the cassette and fluid change process.
3.
When prompted by the controller, disconnect the luer from the Impella
catheter.
Note on Connecting Purge Tubing to the Catheter:
If
included, remove and discard the y-connector and connect the yellow
luer on the purge tubing directly to the yellow check valve on the
Impella Catheter.
4.
Open the Purge Cassette door by pressing the button on the left side
of the console. Remove and discard the old cassette and purge fluid
bag.
5.
Open the new Purge Cassette. Spike the new purge fluid bag with the
new purge cassette tubing. Select
NEXT
to continue.
6.
Insert the new purge cassette into the controller. Be sure to slide the
purge disc into place and extend the purge tubing through the gap in
the purge cassette door when you close the door.
E
NG
LI
SH
Summary of Contents for Impella LD
Page 15: ...13 Impella LD Circulatory Support System ENGLISH...
Page 16: ...Gebrauchsanweisung 14...
Page 30: ...28 Manual del usuario...
Page 44: ...42 Manuale d uso...
Page 58: ...56 Manuale d uso...
Page 72: ...Manuel d utilisation 70...
Page 85: ...83 Impella LD systemet f r cirkulationsst d SVENSKA...
Page 98: ...96 Impella LD kredsl bsst ttesystem DANSK...
Page 111: ...109 Impella LD circulatieodersteunend systeem NEDERLANDS...
Page 112: ...Brukermanual 110...
Page 136: ...134 Instrukcja obs ugi...
Page 149: ...147 System wspomagania kr enia Impella LD POLSKI...
Page 150: ...K ytt ohje 148...
Page 162: ...Manual do Utilizador 160...
Page 175: ...173 Sistema de Suporte Circulat rio Impella LD PORTUGU S...