
PANNORAMIC® CONFOCAL RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
88
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
o
Manual threshold
Set value manually or calculate value by using the
Calculate Threshold
function (can be activated only when manual thresholding is selected). This
sets the sensitivity of the specimen detection. To set the desired threshold
value manually, change the setting in the
Specimen threshold level
selection
box in the range of 0-255. The default value is 212. Smaller values result in a
more sensitive scanning, meaning that more areas will be digitized, however,
this may also include specks or other imperfections of the slide. Greater values
result in a less sensitive scanning, but it is possible that some of the smaller
specimen areas will not be digitized.
Use marker with threshold
If the specimen cannot be identified as required for digitization, (for example, weak contrast), it
can help to encircle completely the specimen with a black marker pen. You have to draw a full
circle with the pen to enable the detection. Marker pen detection will make specimen detection
easier by limiting the scanned area.
•
Manual setting of Use marker with threshold
If this option is selected, only the area encircled with a marker pen will be
scanned. You can define the threshold level by changing the threshold value in the
Marker threshold level
selection box in the range of 0-255. This sets the sensitivity
of the specimen detection inside the marker area. To set the desired threshold
value manually, change the setting in the
Threshold
selection box in the range of
0-255. The default value is 212.
Everything within marker
Available only if
Use marker with threshold
function is activated.
To scan everything within the
marked area, check the
Scan everything within marker
selection box.
Figure 73: Scan everything within the marker selected / not selected