
PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
143
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Virtual slides show poor stitching in Pannoramic® Viewer.
•
Scan camera is not properly adjusted (for example, due to loose fixture).
o
Contact 3DHISTECH Service for support.
Slide is often positioned incorrectly.
•
Guiding fault in transport system
o
Contact 3DHISTECH Service for support.
Digitization process is extremely slow.
•
Extended
focusing
option is activated.
o
Deactivate the
Extended focusing
option if not required.
•
Z-
stack
option is activated.
o
Deactivate the
Z-stack
option if not required.
Image is out of focus (in Manual mode)
•
Focus range limit is activated.
o
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect.
o
Change focus range setting.
All individual images show a shading effect.
•
Algorithm failed to find empty image fields.
o
Capture a reference image with a blank slide.
9.4.1
Fluorescent applications
The switching of the Lumencor cannot be triggered, despite trigger connection through the
camera.
•
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
•
Contact 3DHISTECH Service for support.