PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
142
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
4.
Remove tray manually. If a jammed tray cannot be removed, contact service
personnel.
5.
Close the service door.
6.
Reconnect power supply cable and control computer to line power supply.
7.
Switch on the power supply.
8.
Switch on the control computer.
9.
Launch control software.
Virtual slide is improperly focused almost across its entire surface.
•
Focus range enabled
o
Disable Limiting focus range function.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
o
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory settings give the most
accurate images.
•
Image is of low intensity or noisy.
o
The scannable depth of tissue is highly based on the light-absorption capacity of
the sample. If scanning a thicker tissue sample, the resulting image may become
noisy due to the characteristics of confocal imaging and the loss of intensity.
•
Brightfield image is blurred when using any of the two objectives.
o
Check the cleanness of the condenser and clean it if needed. For more
information, see section
9.5.7 Removing and cleaning the condenser
.
Warning!
Prior performing this operation, the device must be powered off.
•
FOVs are of different intensity on the confocal image.
o
Recalibration must be performed. Contact 3DHISTECH Service for support.