PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
146
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
5.
Replace fuses, put the insert back into its seating, and push it inwards until you hear a
snap, so the lever locks.
Figure 131: Replacing the fuses
9.5.3
Removing and refilling the pipette
The pipette is capable of storing distilled water as immersion fluid for the confocal objective. If
the remaining quantity in the pipette is too low, the software instructs you to remove the pipette
unit and refill with distilled water. Make sure that only premium quality distilled water
(conductance is less than 10 µS/cm) is used.
NOTE:
Two types of pipette that can be used
–
Type 1
and
Type 2
, according to the purchase
order. Type 2 pipette can contain a higher amount of distilled water.
Follow the below steps to remove and refill the pipette if necessary.
1.
Open the pipette door by pushing it at its lower part while the device is powered. The
door lifts automatically.
2.
The pipette is located at the front of the device behind the door (marked with red on the
below image).