PANNORAMIC® CONFOCAL RX 1.0
INSTALLATION
MAY 13, 2021
22
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
2
Installation
Warning!
The installation of Pannoramic® Confocal RX must be carried out by a trained professional!
In any other case the warranty will be automatically voided. The supplier or distributor is not
liable for any damages caused by an installation carried out by an unauthorized person.
Make sure that the unit is positioned in a way that it would be easy to have it disconnected if
necessary.
Warning!
The device of Pannoramic® Confocal RX must be operated on a leveled and solid desk that
can support more weight than of the device's. The calibration of the device must be
performed when it has been installed on the desk. In case the device has been moved to
another location, recalibration must be performed by a qualified and experienced
technician.
Only qualified service assistants are allowed to install the hardware and software. The product is
ready for use after installation.
Caution!
If the temperature of the location where the scanner is operated is much lower than the
temperature of the equipment (for example, the device is brought in from an outer storage,
or has been transported to the site), please leave enough time to normalize the
temperature differences before operating the device within the safe range.
2.1
Operation system of the control computer
The scanner is operable with a Fujitsu CELSIUS R970n as the control computer that runs Windows
10 (see table
Control computer, minimum system requirements
in
).