PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
141
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
9.3.4
Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storag e
temperature above 60°C (140°F) or below 0°C (32°F), which may damage the
equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less
than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
•
Replace DI water in the pipette every week to prevent the reduction of image
quality and fungal infections.
9.4
Troubleshooting
This section describes the most common troubleshooting solutions for the Pannoramic® Confocal
RX product. If any problem arises that does not have solution in this manual, contact the service
personnel.
The product does not start.
•
Power supply is interrupted.
1.
Make sure that the power switch is on (state - 1).
2.
Check cable connections.
Pannoramic® Scanner Software for Research control software has stopped.
•
Slide jammed
1.
Exit control software.
2.
Switch off control computer and the power supply, disconnect power cable.
3.
Open the service door. Follow the steps in
chapter 9.5.1 Opening the service door
.
4.
Remove slide manually (see
chapter 9.5.4 Removing a slide
).
5.
Close the service door.
6.
Reconnect power supply cable and control computer to line power supply.
7.
Switch on the power supply.
8.
Switch on the control computer.
9.
Launch control software.
•
Tray jammed
1.
Exit control software.
2.
Switch off control computer and power supply, disconnect power cable.
3.
Open the service door (see
chapter 9.5.1 Opening the service door
.