PANNORAMIC® CONFOCAL RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
107
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
The following information are included in a profile card:
•
Profile name in the top left corner.
•
Profile status is marked with spots of different color:
o
–
Stored
o
–
Edited
•
Scanning mode icon representing Brightfield or Fluorescence mode
•
Objective type
•
Filter channel color bars (for FL mode only)
•
Scan progress
•
No. of scanned FOVs
•
Scan duration
•
Slide: actual status of the slide is presented here. If the slide is under scanning, the
“Scanning”, if scanning has been interrupted by some error, the “Failed” text is
visible.
NOTE:
If slide status is marked as “Failed”, delete status before scanning by
clicking .
If the virtual slide is created (the scanning is finished), the location path of the
destination folder is displayed. To re-scan and thus recreate the slide, click at
the front of the displayed path to start the scanning process over.
•
Priority scan
•
Scan properties
•
Profile editor for Scheduler
•
Notes
NOTE:
Once scanning has finished, the thumbnail of the digital slide is added to
the card at the bottom. Double click the thumbnail image of the digital slide to
open it in Viewer. If scanning is performed to a local destination, the digital slide
can be opened in CaseViewer and SlideViewer only, but when scanned to
CaseCenter and SlideCenter, it can be opened in InstantViewer as well (depending
on the settings made in
Settings
/
Preferences
/
Server scan result viewer
).