PANNORAMIC® CONFOCAL RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
62
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Preview Type
setting determines the type of illumination. In Brightfield mode when using
Confocal RX, the following preview types are available:
•
Brightfield preview
–
a normal type of illumination in the microscope for
scanning the slide. This option is recommended when scanning in Brightfield
mode.
•
Darkfield preview
–
preview image is created with transmitted Darkfield
illumination. Make sure to select this option in Fluorescent mode.
Remove Coverslip
–
If selected, the edges of the coverslip are not detected as specimen
to be scanned. Depending on the threshold level, it is possible that the edges of the coverslip are
detected as areas to be digitized.
Figure 41: Preview image with and without coverslip edge
Caution!
It requires lots of time (since it cannot find focus neither on the edges, nor on specks) to
digitize coverslip edges. Select
Remove coverslip
function to eliminate those areas.