PANNORAMIC® CONFOCAL RX 1.0
TROUBLESHOOTING AND MAINTENANCE
MAY 13, 2021
145
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Danger of crushing!
Be careful when closing the service door as your fingers may get crushed at the areas of
door hinges.
9.5.2
Replacing fuses of the main power switch
1.
Switch off the main power supply and disconnect the power supply cable (see section
2.
The fuse compartment is located above the main power supply switch at the back side of
the Pannoramic® Confocal RX.
3.
Release lock (
1
) by pushing the lever upwards, so the insert (
2
) can be pulled out.
Figure 129: Fuse compartment
4.
Remove insert from the compartment.
Figure 130: Removing the fuse insert