Do not use excessive force in this step! If you are having trouble inserting the cube it is most
likely due to initial misalignment of the cube with respect to the wheel. Take the cube back
out and try again.
5.
Insert the filter cube in place by pushing it against the filter wheel.
Figure 137: Positioning the filter cube
6.
Check that the filter cube is flush with the wheel.
Warning!
Failure to insert the filter cube properly could result in damage to the instrument and/or
filter cube, as well as compromising the optical performance of the instrument.
7.
Repeat from step
2.
to insert the next filter cube.
8.
When all filter cubes have been inserted close the filter turret door. The instrument will
undergo its initialization procedure during which the wheel spins slowly to detect any
misplaced filter cube or obstruction. If the initialization procedure fails, the filter turret
door indicator will remain dark. If the initialization procedure pass, then the door indicator
will be purple.