PANNORAMIC® CONFOCAL RX 1.0
SCANNING MODES AND SETTINGS
MAY 13, 2021
87
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES .
Caution!
It requires lots of time (since it cannot find focus neither on the edges, nor on specks) to
digitize coverslip edges. Select
Remove coverslip
function to eliminate those areas.
Warning!
Do
not
use
Remove coverslip
function on TMA slides! The layout of TMA slides may mislead
the detecting algorithm.
Fill holes in scanmap
If selected, the holes inside the specimen will also be scanned. Often these holes contain small
sample objects that otherwise would not be scanned.
Remove specks
If selected, the scanner does not digitize objects smaller than the predetermined size. This helps
filtering specks and other uninteresting objects out.
Set the maximum volume of specks manually in the setting window
.
Threshold settings
•
Auto threshold
Select this option to automatically detect the specimen in the preview image.
Important!
Make sure that the detection result is acceptable for you. Otherwise set the threshold value
manually.