G-2
Scan Stitching
Figure G-1: Dektak 150 Stitching Process
M1 = Scan 1
M2 = Scan 2
M1 = [scan length]/2
M2 = {[Scan Length]/2 + Overlap}
Overlap = Scan Length x Overlap Length %
Example:
120 mm Scan length, 10% Overlap
M1 = 60mm, Overlap = 12mm, M2 = 72mm
M2 Start Position is {[scan length]/2 - Overlap} = 48
MODIFYING THE DEFAULT DATA STITCHING
SETTINGS
NOTE –
The Dektak 150 system applies data stitching using the default settings unless you
customize those settings as described below.
1
From the
Setup
menu, select
Configuration Settings
.
2
In the
Configuration Settings
dialog box, click the
Data Stitching
icon. The
Data Stitching
dialog box
appears.