9-10
Calibration and Maintenance
Figure 9-10: Stylus Exchange Tool Details
b.
Place the stylus exchange tool underneath the scan head.
c.
Align the pins of the stylus exchange tool with the stylus sensor housing.
Figure 9-11: Aligning the Stylus Replacement Fixture
d.
Push the stylus exchange tool up until the bottom of the scan head is against the top of it. Be sure the
stylus is seated in the channel on the tool. If necessary, gently tap the end of the stylus shaft (NOT the
tip) with your finger to align the shaft with the channel.
Figure 9-12: Scan Head Flush with Fixture
NOTE –
If the stylus is not properly seated in the channel, it will tend to rotate when you
remove it.
e.
Rotate the magnet wheel to engage the magnet (so that the green dot is visible) in the stylus exchange
tool (see the right side of
).
Magnet Disengaged
(Stylus Released)
Magnet
Wheel,
Green Dot
Visible
Magnet
Wheel,
Notch
Visible
Magnet Engaged
(Stylus Captured)
Alignment
Pins
Stylus
Channel
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......