9-12
Calibration and Maintenance
5
Turn the knob on the stylus exchange tool so that the green dot is visible (right side of
) and
the magnet is engaged.
6
Holding the stylus exchange tool from the edges, align its pins with the outside of the sensor. (Because this
is a precision alignment, you may want to practice several times without the stylus in the way.) Push up
until the tool is flush with the bottom of the sensor (see
Figure 9-15: Aligning the Exchange Tool with the Sensor
7
Turn the thumb wheel to disengage the exchange tool's magnet.
8
Carefully lower the exchange tool to remove it.
9
Figure 9-16: Stylus Shield
10
Place the shield around the sensor head, such that the right-front corner tab is engaged (see
11
Press in on the left -rear corner tab until the shield clicks into place (see
).
Ground clip
Left-rear corner
tab
Right- front
corner tab
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......