6-20
Analytical Functions
Figure 6-20: Smoothing Tab of the Data Processing Dialog Box
3
Choose the desired smoothing
Degree
(1, 2, or 3) on the
Smoothing
tab
4
Determine the smoothing band value by performing the
Total Indicated Reading (TIR)
analytical function
on the scan to be smoothed. Follow the instructions in
Activating the Smoothing Function on page 6-
5
In the
Band
field, enter a value that is equal to or greater than the TIR value shown in the
Analytical
Results
area of the
Data Plot
window.
6
Click
OK
to automatically smooth the profile data whenever the current scan routine executes.
7
To clear smoothing, enter zero (0) in the
Band
field of the
Smoothing
tab of the
Data Processing
dialog
box, and then click
OK
.
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......