Facilities Specifications
A-1
A
F
ACILITIES
S
PECIFICATIONS
As a a high-precision measuring instrument capable of measuring minute physical surface variations,
the Dektak 150 is extremely sensitive to the environment in which it operates. You must set up one of
the following operating environments:
•
Normal Operating Conditions
: The area that houses the Dektak 150 must be free from
excessive dust. Vibration levels must be minimal. The scan head must be protected to eliminate
drafts. The door to the environmental enclosure should be closed both when and when not
making measurements.
•
Reference Operating Conditions
: For very critical measurements, you can purchase an
optional vibration isolation table or platform that is designed for use with the Dektak 150. The
scanning mechanism is sensitive to transient convective flow. After turning on the system,
allow it to stabilize for at least 15 minutes.
lists the facilities requirements of the Dektak 150 system.
lists its dimensions
and weight.
Table A-1: Facilities Requirements
Facility Requirement
Temperature
Operating Range, 20°-25°C (68°-77°F).
Clean Room
Not required. (Class 1000 or better recommended.)
Relative Humidity
50%, +/-20% relative humidity (non-condensing)
Input Voltage
100 - 240 V
~
, 50 - 60 Hz, 3.0 A
Power Demand
720 VA maximum
Power Connection
Four 6-ft, 3-conductor, 16AWG power cords supplied with
system. Cords terminate with male NEMA L5-15
connectors. Connectors rated for 13 amps, 1,625 watts @
125 VAC.
Warm-up Time
15 min. for maximum stability
Vibration
Not to exceed 70
µ
g from 1 to 100 Hz on floor with flat
noise spectrum.
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......