Multiple Scan Operation
5-25
Figure 5-40: Automation Program Options General Tab
2
In the
Scan Data
), do one of the following:
•
Accept the default file to save the scan data.
NOTE –
Unless otherwise specified, data automatically saves to the Default data file in the
DektakV9\Data\Defaul\Default folder on the C: drive.
•
Click the
button on the right to open the
Select Data File
dialog box (see
), where you
can select a data file or specify a new one. For this exercise, enter
Exercise
in the file name field. Then
click
Select
to close the dialog box. Leave the other options at their default settings.
Figure 5-41: Scan Data Saved as a Dataset File
•
Select the
Select At Run Time
check box to wait until the scan has run before you choose the file to save
the scan data.
3
In the
(Scan Data) Export
do one of the following:
•
Select the
None
check box, which prevents ASCII scan data from being exported.
•
Select the
Select At Run Time
check box to wait until the completion of a scan before choosing the file to
save the exported ASCII scan data.
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......