![Stryker Flosteady Arthroscopy Pump 200 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/stryker/flosteady-arthroscopy-pump-200/flosteady-arthroscopy-pump-200_manual_1385909039.webp)
Care and Maintenance
33
EN
CAUTION!
Drying time depends on several variables, including: altitude, humidity, type of
wrap, preconditioning, size of chamber, mass of load, and placement in cham-
ber. Users must verify that drying time set in their autoclave yields dry surgical
equipment.
CAUTION!
Gamma Ray Sterilization
Do not perform a gamma ray sterilization.
*Note
The manufacturer is liable for product defects for the duration of three years in-
sofar they already existed at the time of the risk transfer. Excluded are failures or
outages due to wear and tear of the manual remote control causes by sterilizing
the unit. However, if the user is able to provide evidence that the number of ster-
ilization cycles amounted to less than 100 at the time of failure, the product is re-
placed free of charge.
11.6
Replacing the Fuse
CAUTION!
Before replacing the fuse, check the values of the fuse to be inserted acc. to chap-
ter 13 "Technical Data".
The fuse may be defective and is in need of replacement if:
• displays and LEDs (if available on your equipment) do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does
not
have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power connection cable from mains socket.
4. The fuse holder is located next to the mains socket.
5. Remove fuse holder as depicted in Fig. 11-1 "Opening the fuse holder".
6.
A
Undo the latch of the fuse holder with a small screwdriver.
7.
B
Remove the fuse holder.
8.
C
Check fuse.
9. Insert a new fuse. Use only the specified type of fuse (see chapter 13 "Techni-
cal Data").
10. Insert the fuse holder until it can be heard snapping into place.
11. Use the power cable to reconnect the shockproof safety socket with the rear
mains socket.
Содержание Flosteady Arthroscopy Pump 200
Страница 6: ......
Страница 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Страница 31: ...User Menu 25 EN User Menu Overview ...
Страница 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Страница 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Страница 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Страница 110: ...54 Index DE V Vor der Operation 31 Vorsichtsmaßnahmen 13 W Wartungsintervall alle zwei Jahre 33 ...
Страница 112: ...FR 18 Garantie et assistance technique 49 19 Annexe 51 19 1 Procès verbal de tests 51 Index alphabétique 52 ...
Страница 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Страница 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Страница 164: ......
Страница 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Страница 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Страница 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Страница 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Страница 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Страница 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Страница 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Страница 271: ......