Example:
STAT:PRES
resets the status registers.
:STATus:QUEStionable:CONDition
<Condition>
Queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument.
The content is not deleted after being read out since it indicates the current hardware
status.
Parameters:
<Condition>
string
Example:
:STATus:QUEStionable:CONDition?
queries the Status:Questionable:Condition register.
:STATus:QUEStionable:ENABle
<Enable>
Sets the bits of the ENABle part of the STATus:QUEStionable register. The enable part
determines which events of the STATus:EVENt part are enabled for the summary bit in
the status byte. These events can be used for a service request.
If a bit in the ENABle part is 1, and the correesponding EVENt bit is true, a positive
transition occurs in the summary bit. This transition is reportet to the next higher level.
Parameters:
<Enable>
string
Example:
STAT:QUES:ENAB 1
Problems when performing an adjustment cause an entry to be
made in the sum bit.
:STATus:QUEStionable[:EVENt]
<Event>
Queries the content of the EVENt part of the
STATus:QUEStionable
register. This
part contains information on the actions performed in the instrument since the last
readout. The content of the EVENt part is deleted after being read out.
Parameters:
<Event>
string
Example:
STAT:QUES:EVEN?
queries the Status:Questionable:Event register.
:STATus:QUEStionable:NTRansition
<Ntransition>
Sets the bits of the NTRansition part of the STATus:QUEStionable register. If a bit is
set, a transition from 1 to 0 in the condition part causes an entry to be made in the
EVENt part of the register.
Parameters:
<Ntransition>
string