Configuring the RF Signal
R&S
®
SMBV100B
293
User Manual 1178.4460.02 ─ 03
Settings
............................................................................................... 294
............................................................................................. 296
................................................................................................296
Amplitude
Sets the level of the RF signal.
The value is offset-free and corresponds to the level at the RF connector.
See
"RF frequency and level display with a downstream instrument"
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
●
The display has been disabled for security reasons.
See:
–
–
The display is disabled when:
–
Chapter 6.10, "Varying the RF Signal in List or Sweep
Remote command:
Note:
The SCPI command
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
sets the level of the "Level" display.
This means, the level containing offset.
Limit
Sets an upper limit for the RF output power.
You can use this value to protect your DUT from damage due to high input power. If
you enter an RF level above this value, the instrument limits the output power to this
specified value, and generates the warning message:
"Pep value greater than the defined limit." However, the level indication in the status
bar is not affected.
The setting is not affected by an instrument preset ([Preset] key or
*RST
) and the
"Save/Recall" function. It is reset only by factory preset.
RF Level Settings