![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 672](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881672.webp)
Remote Control Commands
R&S
®
SMBV100B
670
User Manual 1178.4460.02 ─ 03
Example:
See
Example "Connecting and configuring external instrument"
Manual operation:
See
:SCONfiguration:EXTernal:DIGital:RF:FREQuency
<Frequency>
:SCONfiguration:EXTernal:IQOutput:RF:FREQuency
<Frequency>
In uncoupled mode, sets the RF frequency of the external instrument.
Parameters:
<Frequency>
float
Range:
100E3 to 3E9
Increment: 0.01
*RST:
1E9
Example:
SCONfiguration:EXTernal:IQOutput:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput:RF:FREQuency 2143E6
// RF frequency of the external instrument is RF = 2.143 GHz
Manual operation:
See
"(Delta) RF Frequency/ RF Level"
:SCONfiguration:EXTernal:DIGital:RF:FREQuency:OFFSet
<FreqOffset>
:SCONfiguration:EXTernal:IQOutput:RF:FREQuency:OFFSet
<FreqOffset>
In coupled mode, offsets the RF frequency of the external instrument with the selected
delta value.
Parameters:
<FreqOffset>
float
Range:
-3E9 to 3E9
Increment: 0.01
*RST:
0
Example:
SOURce1:FREQuency:CW 2143E6
SCONfiguration:EXTernal:IQOutput:RF:COUPling ON
SCONfiguration:EXTernal:IQOutput:RF:FREQuency:OFFSet 20E6
// The resulting RF frequency of the external instrument is
// RF = 20E6 = 2.163 GHz
// Where both the RF frequency and the frequency offset
// are applied at the external instrument
Manual operation:
See
"(Delta) RF Frequency/ RF Level"
:SCONfiguration:EXTernal:DIGital:RF:POWer
<Power>
:SCONfiguration:EXTernal:IQOutput:RF:POWer
<Power>
In uncoupled mode, sets the RF level of the external instrument.
SCONfiguration Subsystem