Remote Control Commands
R&S
®
SMBV100B
696
User Manual 1178.4460.02 ─ 03
// Configure the frequency modulation settings and switch FM on
SOURce1:FM1:SOURce EXT1
SOURce1:FM1:DEViation 5000
// Query the input sensitivity at the external modulation input
:SOURCe1:FM1:SENSitivity?
// Response: 1000
// since the voltage value for full modulation is 1V,
// the resulting sensitivity is precisely 5000 Hz/V.
// Switch on FM and RF signal output
SOURce1:FM1:STATe 1
OUTPut1:STATe 1
The following commands are available:
......................................................................................696
[:SOURce<hw>]:FM<ch>[:DEViation]
............................................................................... 696
.................................................................................. 697
[:SOURce<hw>]:FM:DEViation:MODE
............................................................................. 697
[:SOURce<hw>]:FM:DEViation:SUM
................................................................................698
..............................................................................................698
.............................................................................................698
.....................................................................................699
[:SOURce<hw>]:FM<ch>:STATe
<State>
Activates frequency modulation.
Suffix:
FM<ch>
.
1..2
determines the modulation signal channel.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
See
Example "Creating a frequency modulated RF signal"
Manual operation:
See
[:SOURce<hw>]:FM<ch>[:DEViation]
<Deviation>
Sets the modulation deviation of the frequency modulation in Hz.
Suffix:
FM<ch>
.
1|2
Modulation signal channel
SOURce Subsystem