Remote Control Commands
R&S
®
SMBV100B
629
User Manual 1178.4460.02 ─ 03
Return values:
<TotalPhysMemKb> integer
Total physical memory.
<ApplicMemKb>
integer
Application memory.
<HeapUsedKb>
integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage:
Query only
12.7
CALibration Subsystem
The
CALibration
subsystem contains the commands needed for performing internal
adjustments. This procedure is triggered by the query commands.
Suffix <hw>
Suffix
Value range
Description
CALibration<hw>
[1]
Optional suffix
Understanding the query response
●
0
: error-free execution of the adjustments
●
1
: indicates that an error occurred; the process has been canceled
......................................................................................... 629
.........................................................................................630
:CALibration<hw>:ALL:INFormation?
............................................................................... 630
........................................................................................ 630
......................................................................................... 631
:CALibration:DATA:FACTory:DATE?
.................................................................................631
:CALibration<hw>:CONTinueonerror
................................................................................ 631
:CALibration<hw>:IQModulator:LOCal?
............................................................................632
:CALibration:ALL[:MEASure]?
[<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Query parameters:
<Force>
string
Return values:
<Measure>
0 | 1 | OFF | ON
CALibration Subsystem