
Configuring the RF Signal
R&S
®
SMBV100B
318
User Manual 1178.4460.02 ─ 03
●
Switches automatically to the next sweep step when the
If the end value is reached, signal generation continues with the next sweep cycle.
●
Stops signal generation with the next external trigger event.
●
Starts the signal generation again with the next trigger event, beginning at the start
value.
●
"State = Off" stops the signal generation in sweep or list mode.
●
Trigger source: An externally applied trigger signal.
Table 6-4: Cross-reference between manual and remote control in Extern Start/Stop modes (sweep)
Manual control mode:
"Extern Start/Stop"
Remote commands
Rohde
&
Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce EAUTo [-]
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
Manual mode (Sweep/List)
The
manual
mode only applies to remote control. It is not visible in the graphical user
interface of the instrument and is described here for completeness.
Figure 6-12: Manual mode in remote control (sweep / list )
●
The instrument generates the signal in steps.
●
Starts signal generation with "State = On".
●
Switches to the next step with a user defined setting via remote control.
Varying the RF Signal in List or Sweep Mode