[:SOURce<hw>]:BBIN:POWer:PEAK
<Peak>
Peak level of the external baseband signal relative to full scale of 0.5 V (in terms of dB
full scale).
Parameters:
<Peak>
float
Range:
-60 to 3.02
Increment: 0.01
*RST:
0
Default unit: dBfs
Example:
See
Example "Applying an external digital baseband signal (Dig
Manual operation:
See
[:SOURce<hw>]:BBIN:POWer:RMS?
Queries the RMS level of the external digital baseband signal.
Return values:
<Rms>
float
Range:
-100 to 10
Increment: 0.01
*RST:
0
Example:
Example "Applying an external digital baseband signal (Dig
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:CDEVice?
Indicates the ID of an externally connected Rohde
&
Schwarz Instrument or
Rohde
&
Schwarz device.
Return values:
<CDevice>
string
"None"
- no device is connected.
Example:
SOURce:BBIN:CDEVice?
Queries the connected device ID.
Usage:
Query only
Manual operation:
See