![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 709](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881709.webp)
Remote Control Commands
R&S
®
SMBV100B
707
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:PULM:DOUBle:DELay
<Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay>
float
*RST:
1E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
Sets the width of the second pulse.
Parameters:
<Width>
float
Increment: 5E-9
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Provided for backward compatibility with former Rohde
&
Schwarz signal generators.
DOUBle
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
[:SOURce<hw>]:PULM:STATe
<State>
Activates pulse modulation.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:SOURce
<Source>
Selects between the internal (pulse generator) or an external pulse signal for the mod-
ulation.