![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 829](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881829.webp)
Remote Control Commands
R&S
®
SMBV100B
827
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:FREQuency:CENTer
<Center>
Sets the center frequency of the sweep.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Center>
float
Range:
300 kHz to RFmax
Increment: 0.01 Hz
*RST:
300E6
Default unit: Hz
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:FREQuency:SPAN
<Span>
Sets the sapn of the frequency sweep range.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Span>
float
Full freqeuncy range
Increment: 0.01
*RST:
400E6
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:FREQuency:STARt
<Start>
Sets the start frequency for the RF sweep.
See
Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Start>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
100 MHz
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
" Start Frequency/Stop Frequency "
SOURce Subsystem