Configuring the RF Signal
R&S
®
SMBV100B
297
User Manual 1178.4460.02 ─ 03
In the "RF Phase" tab, you can determine the delta phase value and reset the
phase to this reference.
The remote commands required to define the settings are described in
ter 12.16.14, "SOURce:PHASe Subsystem"
Settings
......................................................................................... 297
Delta Phase
Sets the phase of the RF signal.
The current phase of the signal is used as the reference.
Note:
Achieving long-term stable phase-locked coupling. We recommend that you use
the LO coupling feature, if you try to synchronize the output signal with the phase of a
second signal generator for example.
See
Chapter 6.9, "Local Oscillator Coupling"
Remote command:
Reset Delta Phase Display
Adopts the selected "Delta Phase" value as the current value, and resets delta phase
to 0 degrees.
Remote command:
RF Phase Settings