Network Operation and Remote Control
R&S
®
SMBV100B
578
User Manual 1178.4460.02 ─ 03
●
For information on how to assign the IP address manually, see
●
Also note that using the RS232 serial interface via USB requires the USB serial
adapter R&S TS-USB1.
11.7
How to Change the GPIB Instrument Address
To control the instrument remotely via the GPIB bus, it must be addressed using the
GPIB address. The remote control address is factory-set to 28, but it can be changed if
it does not fit in the network environment. For remote control, addresses 0 through 30
are allowed. The GPIB address is maintained after a reset of the instrument settings.
To set the GPIB address:
1. Select "System Config > Remote Access > GPIB Address".
2. Select "GPIB Channel Address" and enter a value between 0 and 30.
Risk of losing remote connection
If the remote access and network settings had been configured to values different to
the default, executing a factory preset via remote control terminates the connection to
the instrument.
11.8
How to Set Up a Remote Control Connection
This section guides you through the steps required to set up remote control connec-
tions of the available interfaces.
●
Chapter 11.8.1, "Establishing a Remote Control Connection over the LXI Browser
Interface"
●
Establishing a Remote Control Connection over LAN Using VXI-11 Protocol
●
Setting Up a Remote Control Connection over LAN Using Socket Communication
●
Setting Up a Remote Control Connection over GPIB
●
Setting Up a Remote Control Connection over USB
The instrument and the controller have to be connected with the suitable cable and
switched on.
A remote control program must open a connection to the instrument, before it can send
commands to and receive device responses from the instrument.
How to Set Up a Remote Control Connection