![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 288](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881288.webp)
Configuring the RF Signal
R&S
®
SMBV100B
286
User Manual 1178.4460.02 ─ 03
6
Configuring the RF Signal
The vector signal generator R&S
SMBV100B generates RF signals with outstanding
spectral purity within the frequency range from 8
kHz up to 6
GHz and with adjustable
signal level over a wide range.
In addition to these real-time CW signals, you can generate RF signals from predefined
lists and sweep signals that vary according to the frequency or amplitude curves.
You can also apply versatile analog modulation types with definable characteristics.
The variably adjustable output level due to the built-in attenuator, allows you to vary the
RF signal level over the full level range. There are different methods to improve signal
performance and to optimize the signal quality for the particular application, or to
increase the accuracy and reliability of the generated RF signal.
The R&S
SMBV100B supports R&S NRP power sensors, e.g. to monitor the output
level of the generator and to determine the level correction values for user correction
lists.
Signal modes and characteristics
The R&S
SMBV100B generates unmodulated or analog modulated RF signals. You
can output the signal in fixed mode or as a signal having periodically varying frequen-
cies or amplitudes.
Signal modes
for RF signal generation:
●
Unmodulated signal
Generates an unmodulated continuous wave (CW) of constant frequency and
amplitude.
For information on the signal frequency and level settings and example on how to
configure a simple CW signal, see:
–
Chapter 6.5, "RF Frequency Settings"
–
Chapter 6.6, "RF Level Settings"
–
Chapter 2.3.1, "Generating an Unmodulated Carrier"
●
Analog modulated signal
Modulates the signal with an analog signal and generates amplitude, phase, fre-
quency and pulse modulation.
See
Chapter 6.11, "Analog Modulations"
●
Vector modulated signal
Generates an I/Q vector modulated signal
See
Chapter 5, "Applying I/Q Vector Modulation"
The R&S
SMBV100B allows you to provide the RF signal with constant or varying fre-
quencies and/or amplitudes at the output:
●
Constant frequency and level (CW/Fixed mode)
The RF output signal has the set frequency and level.
●
Varying frequency and/or level
–
Sweep mode