Multi-Instrument Setups
R&S
®
SMBV100B
462
User Manual 1178.4460.02 ─ 03
Remote command:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe
Measurement Period
Sets the recording duration for measuring the baseband input signal by
"Auto
Level
Set".
Note:
For accurate level measurements, set the measurement period to a time value
that is long enough to capture several periods of the input signal.
Remote command:
Auto Level Set
Starts measuring the input signal for estimating the crest factor, the peak power, and
the RMS power of the input signal.
See also
"Fulfilling the prerequisites for a correct signal processing of the externally
supplied digital baseband signal"
Remote command:
[:SOURce<hw>]:BBIN:ALEVel:EXECute
8.6.3
Sample Rate Settings
Access:
1. Select "BB Input > Digital I/Q In > Baseband Input Settings > Interface = DIG I/Q ".
2. Select "Sample Rate".
Sample Rate Source
Indicates the digital interface used to estimate the sample rate.
"Digital I/Q In"
Estimates the sample rate based on the applied I/Q data clock and
displays the resulting value in the value field. The I/Q data clock sig-
nal is fed via the Dig I/Q input.
Note:
Estimation works best if the sample rate is close to the user-
defined value.
Digital Baseband Input Settings