![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 682](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881682.webp)
Remote Control Commands
R&S
®
SMBV100B
680
User Manual 1178.4460.02 ─ 03
Usage:
Event
Manual operation:
See
[:SOURce]:INPut:USER:CLOCk:LEVel
<Level>
[:SOURce]:INPut:USER:TRIGger:LEVel
<Level>
Sets the threshold for any input signal at the User1-2 connectors.
Parameters:
<Level>
float
Range:
0.1 to 2
Increment: 0.1
*RST:
1
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
[:SOURce]:INPut:USER:PULM:LEVel
<Level>
Sets the threshold for any input signal at the User3-5 connectors.
Parameters:
<Level>
float
Range:
0.1 to 2
Increment: 0.1
*RST:
1
Default unit: V
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
"Threshold User3-5 Input/Threshold Pulse Input"
[:SOURce]:INPut:USER:CLOCk:IMPedance
<Impedance>
[:SOURce]:INPut:USER:TRIGger:IMPedance
<Impedance>
Selects the input impedance for the external trigger inputs.
Parameters:
<Impedance>
G1K | G50
*RST:
G1K
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
"Impedance Clock/Trigger Input"
SOURce Subsystem