Remote Control Commands
R&S
®
SMBV100B
871
User Manual 1178.4460.02 ─ 03
Parameters:
<Amplitude>
float
Range:
depends on the installed options
Increment: 0.01
*RST:
n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation:
See
[:SOURce<hw>]:POWer:LMODe
<LevMode>
Sets the RF level mode.
Parameters:
<LevMode>
NORMal | LOWNoise | LOWDistortion
NORMal
Supplies the RF signal with the standard power level of the
instrument.
LOWNoise
Supplies a very low noise sinewave signal.
LOWDistortion
Supplies a very pure sinewave signal.
*RST:
NORMal
Example:
SOURce1:POWer:LMODe LOWD
Sets low distortion mode. The instrument reduces distortions of
the RF signal to a minimum.
Manual operation:
See
[:SOURce<hw>]:POWer:MANual
<Manual>
Sets the level for the subsequent sweep step if
.
Use a separate command for each sweep step.
Parameters:
<Manual>
float
You can select any level within the setting range, where:
STARt
STOP
OFFSet
Range:
(STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
Example:
See
Example "Setup an RF frequency or power sweep"