Remote Control Commands
R&S
®
SMBV100B
826
User Manual 1178.4460.02 ─ 03
Range:
(STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01Hz
*RST:
100 MHz
Default unit: Hz
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:FREQuency:MULTiplier
<Multiplier>
Sets the multiplication factor N
FREQ:MULT
of a subsequent downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
.
The query
returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See
"RF frequency and level display with a downstream instrument"
Parameters:
<Multiplier>
float
Range:
1 to dynamic
Increment: 0.001
*RST:
1
Example:
See
Example "Frequency configuration"
Manual operation:
See
[:SOURce<hw>]:FREQuency:OFFSet
<Offset>
Sets the frequency offset f
FREQ:OFFSet
of a downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
.
The query
returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See
"RF frequency and level display with a downstream instrument"
Note:
The offset also affects RF frequency sweep.
Parameters:
<Offset>
float
Increment: 0.01
*RST:
0
Example:
See
Example "Frequency configuration"
Manual operation:
See
SOURce Subsystem