![Rohde & Schwarz SMBV100B Скачать руководство пользователя страница 695](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881695.webp)
Remote Control Commands
R&S
®
SMBV100B
693
User Manual 1178.4460.02 ─ 03
Suffix:
<ch>
.
1...2
Modulation signal channel
Parameters:
<Depth>
float
Range:
0 to 100
Increment: 0.01
*RST:
30
Example:
See
Example "Creating an amplitude modulated RF signal"
Manual operation:
See
[:SOURce<hw>]:AM:DEPTh:SUM
<AmDepthSum>
Sets the total depth of the LF signal when using combined signal sources in amplitude
modulation.
Parameters:
<AmDepthSum>
float
Range:
0 to 100
Increment: 0.01
*RST:
30
Example:
See
Example "Using combined LF signals"
Manual operation:
See
[:SOURce<hw>]:AM:DEViation:MODE
<AmDevMode>
Selects the coupling mode. The coupling mode parameter also determines the mode
for fixing the total depth.
Parameters:
<AmDevMode>
UNCoupled | TOTal | RATio
UNCoupled
Does not couple the LF signals.
The deviation depth values of both paths are independent.
TOTal
Couples the deviation depth of both paths.
RATio
Couples the deviation depth ratio of both paths
*RST:
UNCoupled
Example:
See
Example "Using combined LF signals"
Manual operation:
See