Remote Control Commands
R&S
®
SMBV100B
615
User Manual 1178.4460.02 ─ 03
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code.
Return values:
<ErrorCode>
integer > 0 (in decimal format)
An error occurred.
(For details, see the Service Manual supplied with the instru-
ment).
0
No errors occurred.
Usage:
Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
12.5
Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section.
In addition, a specific preset command is provided for each digital standard. These
specific commands are described in the associated subsystems.
Four presetting actions are available:
●
Activating the default state of all internal instrument functions (
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
●
Activating the preset state of the parameters related to the selected signal path (
●
Activating the preset state of all parameters that are not related to the signal path
(
●
Activating the original state of delivery (factory reset,
on page 616). Only functions that are protected by a password remain
unchanged as well as the passwords themselves.
Preset Commands