Remote Control Commands
R&S
®
SMBV100B
706
User Manual 1178.4460.02 ─ 03
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate
If
[:SOURce<hw>]:PULM:TRIGger:MODE
SINGle
, triggers the pulse generator.
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:PERiod
<Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period>
float
The minimum value depends on the installed options
R&S
SMBVB-K22 or R&S
SMBVB-K23
Range:
20E-9 to 100
Increment: 5E-9
*RST:
10E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width>
float
Range:
20E-9 to 100
Increment: 10E-9
*RST:
2E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:DELay
<Delay>
Sets the pulse delay.
Parameters:
<Delay>
float
*RST:
1ms
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
SOURce Subsystem