Configuring the Internal Baseband Source
R&S
®
SMBV100B
134
User Manual 1178.4460.02 ─ 03
With long control lists, it is helpful to display only a part of the control list. In such
cases, set the "Visible/Bits Visible" to determine the number of symbols/bits to be dis-
played and select "Zoom" to focus the displayed area around the current "Cursor Posi-
tion".
Ramps outside the displayed area are not lost by zooming.
Remote command:
n.a.
Save/Save As
Stores the changes in the selected control list file or in a new file.
Remote command:
n.a.
3.5.4
How to Create Data and Control Lists
This section provides step-by-step instructions on configuring and using the provided
settings. For details on individual functions and settings, see
3.5.4.1
How to Create and Assign a Control List
The R&S
SMBV100B provides the following ways to create a file containing control sig-
nals:
●
Control and Marker Lists Editor
and create a file in ASCII for-
mat and with extension
*.dm_iqc
, see
"To create a control list in ASCII format
●
To use the tag-oriented format and create a control list file, see
●
To use SCPI commands and create a file in binary format, see
To create a control list in ASCII format manually
Use the intuitive build in
Control and Marker Lists Editor
dialog:
1. To open the "Control Data Editor":
a) Select "Baseband > Custom Digital Mod > Data Source > Select Control List".
b) Select an existing file or select "New".
If the selected file is an existing one, the dialog shows the used settings.
If a new file has been created, the control list is empty.
2. Adjust the control signals as required.
a) In the graphic editor "Select Ramp to Edit", select the color-coded trace of the
required signal.
b) Tap at a position which requires a ramp. The number or ramps per marker is
unlimited.
Generating Custom Digital Modulated Signals