Remote Control Commands
R&S
®
SMBV100B
661
User Manual 1178.4460.02 ─ 03
Note:
When measuring the RF signal, the sensor considers the corresponding correc-
tion factor at that frequency, and uses the level setting of the instrument as reference
level.
Parameters:
<Source>
A | USER | RF
*RST:
A
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:STATus[:DEVice]?
Queries if a sensor is connected to the instrument.
Return values:
<Status>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS1:STAT?
Response: 1
A sensor is connected.
Usage:
Query only
Manual operation:
See
:SENSe<ch>[:POWer]:SVERsion?
Queries the software version of the connected R&S NRP power sensor.
Return values:
<SVersion>
string
Example:
SENS1:POW:SVER?
Queries the software version of the power sensor.
Usage:
Query only
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type>
string
Example:
SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage:
Query only
SENSe, READ, INITiate and SLISt Subsystems