Waveform setup
R&S
®
RTA4000
75
User Manual 1335.7898.02 ─ 08
"En
HR"
Each acquisition is done in high resolution mode, and the minimum
and maximum values over some consecutive acquisitions build the
envelope.
Remote command:
No. of Averages
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
To restart the average calculation, press the [Clear Screen] key.
Remote command:
Nx Single
Sets the number of waveforms that are acquired with a [Single] acquisition.
Remote command:
Roll
Enables the automatic roll mode. The instrument switches to roll mode if the
The roll mode displays the untriggered, continuous signal, and moves the captured
input data on the display from the left to the right. The instrument shows the waveform
immediately, without waiting for the complete acquisition of the waveform record. The
record length is set automatically ("Auto" mode). Some math functions are non-calcula-
ble if roll mode is active.
You can use the horizontal and vertical zoom in roll mode if the acquisition is stopped.
Remote command:
Start Roll Time
Sets the limit timebase for the roll mode. The instrument switches automatically to roll
mode if:
●
the
exceeds the value given here.
●
the roll mode is activated (
Remote command:
Interpolation
Selects the interpolation method if interpolation is required to get the defined record
length.
Acquisition setup